Team:METU Turkey/Results Discussion/PCR

From 2010.igem.org

(Difference between revisions)
(New page: <br>MAIN STEPS/ TIME TABLE <br>- Each step is carried out for 1 tube <br>- PCR solution preparation [30-45 min] <br>- Thermocycler [2.5 h] <br>- Denaturation <br>- Annealing <b...)
 
(One intermediate revision not shown)
Line 1: Line 1:
-
<br>MAIN STEPS/ TIME TABLE
+
<br><h3>MAIN STEPS/ TIME TABLE</h3>
<br>- Each step is carried out for 1 tube
<br>- Each step is carried out for 1 tube
<br>- PCR solution preparation [30-45 min]
<br>- PCR solution preparation [30-45 min]
Line 11: Line 11:
-
<br>MATERIALS       
+
<br><h3>MATERIALS</h3>        
<br>- Thermal Cycler [M08-Techne TC-512 Gradient PCR]
<br>- Thermal Cycler [M08-Techne TC-512 Gradient PCR]
<br>- Spin [M03]
<br>- Spin [M03]
Line 27: Line 27:
<br>- PCR purification kit
<br>- PCR purification kit
-
<br>SOLUTIONS
+
<br><h3>SOLUTIONS</h3>
<br>- PCR MasterMix Kit [Fermentas]
<br>- PCR MasterMix Kit [Fermentas]
        
        
-
<br>CHECK-LIST PROCEDURE
+
<br><h3>CHECK-LIST PROCEDURE</h3>
<br>Sample Preparation
<br>Sample Preparation
<br>- Calculate the amount of your samples.
<br>- Calculate the amount of your samples.
Line 48: Line 48:
<br>- Step 7 (store): 4 C, forever.
<br>- Step 7 (store): 4 C, forever.
-
<br>Follow-Up Steps
+
<br><h3>Follow-Up Steps</h3>
<br>(It is recommended that you use PCR products for RE digestion and ligation in the same day to prevent the damage to DNA ends)
<br>(It is recommended that you use PCR products for RE digestion and ligation in the same day to prevent the damage to DNA ends)
Line 55: Line 55:
-
<br>NOTES
+
<br><h3>NOTES</h3>
<br>- To reduce the chance of contamination with exogenous DNAs, bake all glassware for 6 h at 15 C and autoclave all plasticware.
<br>- To reduce the chance of contamination with exogenous DNAs, bake all glassware for 6 h at 15 C and autoclave all plasticware.
<br>- If the thermal cycler is not equipped with a heated lid, use either mineral oil or paraffin wax to prevent evaporation of liquid from the reaction mixture.
<br>- If the thermal cycler is not equipped with a heated lid, use either mineral oil or paraffin wax to prevent evaporation of liquid from the reaction mixture.
Line 63: Line 63:
   
   
-
<br>TROUBLESHOOTING
+
<br><h3>TROUBLESHOOTING</h3>
<br>No bands
<br>No bands
<br>- PCR kit or polymerase enzyme is not working: Include positive control (ex. a control DNA sequence with primers)
<br>- PCR kit or polymerase enzyme is not working: Include positive control (ex. a control DNA sequence with primers)
Line 69: Line 69:
     <br>- Secondary structure: Use DMSO for helping denaturation of DNA secondary structures
     <br>- Secondary structure: Use DMSO for helping denaturation of DNA secondary structures
     <br>- Melting temperatures are different: see annealing temperature  
     <br>- Melting temperatures are different: see annealing temperature  
-
     <br>- Annealing temperature is not right: Use 50 C. It is used as a generic temperature by sequencing center to amplify all different samples.
+
     <br>- Annealing temperature is not right: Use 50 C. It is used as a generic temperature by sequencing center to <br>amplify all different samples.
<br>- Problem with template and primer working solutions:   
<br>- Problem with template and primer working solutions:   
     <br>- Measure concentration.  
     <br>- Measure concentration.  
Line 79: Line 79:
-
<br>Alternative Procedure  
+
<br><h3>Alternative Procedure</h3>
<br>- Master Mix
<br>- Master Mix
     <br>- 3 uL-10X Taq Buffer
     <br>- 3 uL-10X Taq Buffer
Line 91: Line 91:
     <br>- 3 uL-250 ng/uL Template
     <br>- 3 uL-250 ng/uL Template
-
<br>Alternative Procedure with Pfu DNA polymerase
+
<br><h3>Alternative Procedure with Pfu DNA polymerase</h3>
<br>- 10 uL 10X Pfu Buffer
<br>- 10 uL 10X Pfu Buffer
<br>- 8 uL dNTPs (2.5 mM stock)
<br>- 8 uL dNTPs (2.5 mM stock)

Latest revision as of 03:33, 28 October 2010


Contents

MAIN STEPS/ TIME TABLE


- Each step is carried out for 1 tube
- PCR solution preparation [30-45 min]
- Thermocycler [2.5 h]

   
- Denaturation
- Annealing
- Primer extension


- PCR purification [30 min]
- DNA concentration determination with Nanodrop [5 min]
- Electrophoresis [2 h]



MATERIALS


- Thermal Cycler [M08-Techne TC-512 Gradient PCR]
- Spin [M03]
- P10 and P100 micropipetes
- DNase-free micropipet tips
- Microfuge tubes (0,5 mL, thin-walled for amplification reactions)
- Electrophoresis [MO6]
- Nanodrop [M08]


- PCR kit (MasterMix)
- Forward primer
- Reverse primer
- Template DNA
- Nuclease-free water
- PCR purification kit


SOLUTIONS


- PCR MasterMix Kit [Fermentas]



CHECK-LIST PROCEDURE


Sample Preparation
- Calculate the amount of your samples.
- Design the tubes.
- Label the microfuge tubes.
- Transfer the PCR reaction to a pre-heated block.


- Adjustment of thermocycler [changeable parameters]
- Lid Temp: 105 C (to prevent evaporation)
- Step 1 (initial denaturation): 94 C, 5 min. Press select to go.
- Step 2 (cycle denaturation): 94 C, 30 sec.
- Step 3 (cycle annealing): primer specific, 30 sec.
- Step 4 (cycle extension): 72 C, 1 min. Press select two times.
- Step 5 (cycle loop): Go to step 2, 30 to 35 cycles.
- Step 6 (final extension): 72 C, 15 min.
- Step 7 (store): 4 C, forever.


Follow-Up Steps


(It is recommended that you use PCR products for RE digestion and ligation in the same day to prevent the damage to DNA ends)


Agarose gel electrophoresis
- PCR product purification is NOT needed for gel loading. You can directly load the PCR product to check the results.



NOTES


- To reduce the chance of contamination with exogenous DNAs, bake all glassware for 6 h at 15 C and autoclave all plasticware.
- If the thermal cycler is not equipped with a heated lid, use either mineral oil or paraffin wax to prevent evaporation of liquid from the reaction mixture.
- High concentrations of of dNTPs (> 4mM) are inhibitory, because of sequestering Mg2+.
- Pfu is more suitable, error rate of it per nucleotide is the lowest of any DNA polymerase.
- Inhibition of PCR can be caused by proteinase K, phenol, EDTA, ionic detergents, heparin, gel-loading dyes (bromophenol blue, xylene cyanol)



TROUBLESHOOTING


No bands
- PCR kit or polymerase enzyme is not working: Include positive control (ex. a control DNA sequence with primers)
- Primers are not working

   
- Secondary structure: Use DMSO for helping denaturation of DNA secondary structures
- Melting temperatures are different: see annealing temperature
- Annealing temperature is not right: Use 50 C. It is used as a generic temperature by sequencing center to
amplify all different samples.


- Problem with template and primer working solutions:

   
- Measure concentration.
- Include positive control (ex. a control DNA sequence with primers)
- Increase the amount of template


Non-specific bands
- If you see no PCR product then decrease annealing temp 5-10 C; if you see non-specific products, then increase it by 5-10 C.



Alternative Procedure


- Master Mix

   
- 3 uL-10X Taq Buffer
- 3 uL-2mM dNTP mix
- 3 uL-25mM MgCl2
- 18 uL Nuclease-free water


- For 30 uL PCR mix

   
- 27 uL Master mix
- 3 uL-5uM forvard primer
- 0.25 uL-5 u/uL Taq
- 3 uL-250 ng/uL Template

Alternative Procedure with Pfu DNA polymerase


- 10 uL 10X Pfu Buffer
- 8 uL dNTPs (2.5 mM stock)
- 2.5 uL primer 1 (100 ng/uL)
- 2.5 uL primer 2 (100 ng/uL)
- 1 uL Pfu enzyme
- x uL template ( around 10 ng)
- Complete to 100 uL with nuclease-free water


- 94 C 1 min
- 53 C 1 min
- 72 C 1 min 30 sec for 30 cycles
- End on a 10 min 72 C extension and keep at 4 C.


- Extension times are longer for Pfu than Taq.
- If amplification does not work for first time, add extra MgCl2.