Preparing 1% Agarose Gels:
- Added 180 mL 1x TBE and 1.8g agarose
- Heated up until agarose was no longer visible
- Let cool (approx. 10 min)
- Added 180 սL Ethidium bromide (1000X)
PCR Purification
- Added 5 volumes of Buffer PBI to 1 volume of the PCR sample and mix (in a clean 1.5 mL eppendorf)
- Checked that the color of the mixture was yellow (similar to Buffer PBI without the PCR sample).
- To bind DNA, transferred the sample to the column and centrifuge at 17,900g for 30 – 60 secs.
- Discarded flow-through. Placed the column back in the same tube.
- To wash, added 0.75 mL Buffer PE to the column and centrifuged for 30 – 60 secs.
- Discarded flow-through and place the column back in the same tube. Centrifuged the column for an additional 1 min.
- Placed the column in a clean 1.5 mL microcentrifuge tube.
- To elute DNA, 50 µL autoclaved milliQ water to the center of the membrane, let the column sit for 1 min, and then centrifuged.
Heat shock transformation of the plasmids into our bacteria
- Left cells and ligation reaction products on ice.
- Added plasmid (5սL) to cells. Mix gently by swirling pipette tip in mixture (DO NOT ASPIRATE).
- Left cells on ice for 30 min.
- Applied heat shock of 45 seconds in 42C bath.
- Put tubes on ice for 2 min.
- Added 250 սL of LB (room temp.)
- Incubated 1 hour at 37 C
- Plated 100սL and left plate in the 37 degrees incubator
- Incubated overnight at 37C.
Gel Extraction
- Excise DNA fragment from the agarose gel with a clean, sharp scalpel.
- Weigh the gel slice in a colorless tube. Add 3 volumes of Buffer QG to 1 volume of gel (100 mg or approximately 100 μL).
- Incubate at 50ºC for 10 min (or until the gel slice has completely dissolved). To help dissolve gel, mix by vortexing the tube every 2 – 3 min during the incubation.
- After the gel slice has completely dissolved, check that the color of the mixture is yellow (similar to Buffer QG without dissolved agarose).
- Add 1 gel volume of isopropanol to the sample and mix.
- Place a QIAquick spin column in a provided 2 mL collection tube.
- To bind DNA, apply the sample to the QIAquick column, and centrifuge for 1 min.
- Discard flow-through and place QIAquick column back in the same collection tube.
- Recommended: Add 0.5 mL of Buffer GQ to QIAquick column and centrifuge for 1 min.
- To wash, add 0.75 mL of Buffer PE to QIAquick column and centrifuge for 1 min.
- Discard the flow-through and centrifuge the QIAquick column for an additional 1 min at 17,900 x g (13,000 rpm).
- Place QIAQuick column into a clean 1.5 mL microcentrifuge tube.
- To elute DNA, add 50 μL of Buffer EB (1 mM Tris-Cl, pH 8.5) or water (pH 7.0 – 8.5) to the center of the QIAquick membrane and centrifuge the column for 1 min. Alternatively, for increased DNA concentration, add 30 μL elution buffer to the center of the QIAquick membrane, let the column stand for 1 min, and then centrifuge for 1 min.
- If the purified DNA is to be analyzed on a gel, add 1 volume of Loading Dye to 5 volumes of purified DNA. Mix the solution by pipetting up and down before loading the gel.
Plasmid DNA Purification with Mini Prep Kit
- Resuspend pelleted bacterial cells in 250 µL Buffer P1 and transfer to a microcentrifuge tube.
- Add 250 µL Buffer P2 and mix thoroughly by inverting the tube 4 - 6 times.
- Add 350 µL Buffer N3 and mix immediately and thoroughly by inverting the tube 4 – 6 times.
- Centrifuge for 10 mins at 13,000 rpm (~17,900 xg) in a table-top microcentrifuge.
- Apply the supernatants from step 4 to the spin column by decanting or pipetting.
- Centrifuge for 30 – 60 secs. Discard the flow-through.
- Recommended: Wash the spin column by adding 0.5 mL Buffer PB and centrifuging for 30 – 60 secs. Discard the flow-through.
- Wash spin column by adding 0.75 mL Buffer PE and centrifuging for 30 – 60 secs.
- Discard the flow-through, and centrifuge for an additional 2 min to remove residual wash buffer.
- Place the column in a clean 1.5 mL microcentrifuge tube. To elute DNA, add 30 µL autoclaved water to the center of each spin column, let stand for 1 min, and centrifuge for 1 min.