Team:METU Turkey/Results Discussion/CS

From 2010.igem.org

(Difference between revisions)
 
Line 356: Line 356:
     </ul><!--topnav ends-->
     </ul><!--topnav ends-->
</div><!--tab_nav ends-->
</div><!--tab_nav ends-->
-
 
+
<br> -- </br>
<br>10/09/26
<br>10/09/26
-
<br>Purification with CS BL21 pTriEx
+
<br><h3>Purification with CS BL21 pTriEx</h3>
<br>Loaded combined FPLC tubes from QS column(10/09/26) gave two close picks in CS column. First one corresponds to the CooA. The elutions of CooA are C13,C14 and C15. We combined them for further analyzes with EMSA and ITC and stored @+4. C16 AND C17 are combined to reload to CS column because they are contaminated another protein which is close to 40kDa.
<br>Loaded combined FPLC tubes from QS column(10/09/26) gave two close picks in CS column. First one corresponds to the CooA. The elutions of CooA are C13,C14 and C15. We combined them for further analyzes with EMSA and ITC and stored @+4. C16 AND C17 are combined to reload to CS column because they are contaminated another protein which is close to 40kDa.
Line 374: Line 374:
<br>10/10/12 CS
<br>10/10/12 CS
-
<br>Purification with CS BL21 pTriEx
+
<br><h3>Purification with CS BL21 pTriEx</h3>
<br>According to SDS-GEL, we combined 4,6,7,8 and 9 tubes containing 0,5ml and concentrated to 1ml for furhter analyzes such EMSA and ITF. We measured absorbance of combined @ 280,260 and 420 as respectively 2.76, 2.83 and 0.837. CooA concentration was determined as 7.75 uM according to extinction coefficient of heme(Ferric State) which is 108 mM-1 cm-1 according to Aono. We didnot take 5th tube because it contains much more impurities than other tubes.
<br>According to SDS-GEL, we combined 4,6,7,8 and 9 tubes containing 0,5ml and concentrated to 1ml for furhter analyzes such EMSA and ITF. We measured absorbance of combined @ 280,260 and 420 as respectively 2.76, 2.83 and 0.837. CooA concentration was determined as 7.75 uM according to extinction coefficient of heme(Ferric State) which is 108 mM-1 cm-1 according to Aono. We didnot take 5th tube because it contains much more impurities than other tubes.

Latest revision as of 03:41, 28 October 2010

Home

 


--

10/09/26

Purification with CS BL21 pTriEx


Loaded combined FPLC tubes from QS column(10/09/26) gave two close picks in CS column. First one corresponds to the CooA. The elutions of CooA are C13,C14 and C15. We combined them for further analyzes with EMSA and ITC and stored @+4. C16 AND C17 are combined to reload to CS column because they are contaminated another protein which is close to 40kDa.
w7


w7



10/10/12 CS


Purification with CS BL21 pTriEx


According to SDS-GEL, we combined 4,6,7,8 and 9 tubes containing 0,5ml and concentrated to 1ml for furhter analyzes such EMSA and ITF. We measured absorbance of combined @ 280,260 and 420 as respectively 2.76, 2.83 and 0.837. CooA concentration was determined as 7.75 uM according to extinction coefficient of heme(Ferric State) which is 108 mM-1 cm-1 according to Aono. We didnot take 5th tube because it contains much more impurities than other tubes.


w7